Phantom Study of an In-House Amplitude-Gating Respiratory Method with Silicon Photomultiplier Technology Positron Emission Tomography/Computed Tomography. - Anthropologie Bio-culturelle, Droit, Éthique, Santé - UMR 7268 ADES
Article Dans Une Revue Computer Methods and Programs in Biomedicine Année : 2022

Phantom Study of an In-House Amplitude-Gating Respiratory Method with Silicon Photomultiplier Technology Positron Emission Tomography/Computed Tomography.

Résumé

PURPOSE: The objective of this phantom study was to determine whether breathing-synchronized, silicon photomultiplier (SiPM)-based PET/CT has a suitable acquisition time for routine clinical use. METHODS: Acquisitions were performed in list mode on a 4-ring SiPM-based PET/CT system. The experimental setup consisted of an external respiratory tracking device placed on a commercial dynamic thorax phantom containing a sphere filled with [F-18]-fluorodeoxyglucose. Three-dimensional sinusoidal motion was imposed on the sphere. Data were processed using frequency binning and amplitude binning (the "DMI" and "OFFLINE" methods, respectively). PET sinograms were reconstructed with a Bayesian penalized likelihood algorithm. RESULTS: Respiratory gating from a 150-sec acquisition was successful. The DMI and OFFLINE methods gave similar activity profiles but both were slightly shifted in space; the latter profile was closest to the reference acquisition. CONCLUSION: With SiPM PET/CT systems, the amplitude-based processing of breathing-synchronized data is likely to be feasible in routine clinical practice.
Fichier principal
Vignette du fichier
S0169260722002899.pdf (1.96 Mo) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-03692953 , version 1 (22-07-2024)

Licence

Identifiants

Citer

Pascal Bailly, Roger Bouzerar, Romain Galan, Marc-Etienne Meyer. Phantom Study of an In-House Amplitude-Gating Respiratory Method with Silicon Photomultiplier Technology Positron Emission Tomography/Computed Tomography.. Computer Methods and Programs in Biomedicine, 2022, 221, pp.106907. ⟨10.1016/j.cmpb.2022.106907⟩. ⟨hal-03692953⟩
46 Consultations
8 Téléchargements

Altmetric

Partager

More