Noise modeling for intensified camera in fluorescence imaging: application to image denoising - Compartimentation et dynamique cellulaires Accéder directement au contenu
Communication Dans Un Congrès Année : 2013

Noise modeling for intensified camera in fluorescence imaging: application to image denoising

Résumé

We propose a statistical framework for noise variance estimation in the case of experimental microscopic images enhanced by an image intensifier. Instrumentally induced noise is modeled and corrected to cope with optical aberrations. In this paper, the spatially varying noise is exploited for denoising applications. Our approach does not need variance stabilization since the algorithm is able to adapt to local noise statistics. Performances are demonstrated on real samples on both widefield and confocal imaging combined with frequency domain fluorescence lifetime imaging (FD FLIM).
Fichier non déposé

Dates et versions

hal-00931360 , version 1 (15-01-2014)

Identifiants

Citer

Philippe Roudot, Charles Kervrann, Jérôme Boulanger, François Waharte. Noise modeling for intensified camera in fluorescence imaging: application to image denoising. ISBI - IEEE International Symposium on Biomedical Imaging, Apr 2013, San-Francisco, United States. pp.600-603, ⟨10.1109/ISBI.2013.6556546⟩. ⟨hal-00931360⟩