Filtered backprojection proton CT reconstruction along most likely paths

S. Rit
N. Freud
D. Sarrut
J.-M. Letang

Résumé

Purpose:Proton CT (pCT) has the potential to accurately measure the electron density map of tissues at low doses but the spatial resolution is prohibitive if the curved paths of protons in matter is not accounted for. The authors propose to account for an estimate of the most likely path of protons in a filtered backprojection (FBP) reconstruction algorithm.Methods:The energy loss of protons is first binned in several proton radiographs at different distances to the proton source to exploit the depth-dependency of the estimate of the most likely path. This process is named the distance-driven binning. A voxel-specific backprojection is then used to select the adequate radiograph in the distance-driven binning in order to propagate in the pCT image the best achievable spatial resolution in proton radiographs. The improvement in spatial resolution is demonstrated using Monte Carlo simulations of resolution phantoms.Results:The spatial resolution in the distance-driven binning depended on the distance of the objects from the source and was optimal in the binned radiograph corresponding to that distance. The spatial resolution in the reconstructed pCT images decreased with the depth in the scanned object but it was always better than previous FBP algorithms assuming straight line paths. In a water cylinder with 20 cm diameter, the observed range of spatial resolutions was 0.7 − 1.6 mm compared to 1.0 − 2.4 mm at best with a straight line path assumption. The improvement was strongly enhanced in shorter 200° scans.Conclusions:Improved spatial resolution was obtained in pCT images with filtered backprojection reconstruction using most likely path estimates of protons. The improvement in spatial resolution combined with the practicality of FBP algorithms compared to iterative reconstruction algorithms makes this new algorithm a candidate of choice for clinical pCT.
Fichier non déposé

Dates et versions

hal-00796949 , version 1 (05-03-2013)

Identifiants

Citer

S. Rit, G. Dedes, N. Freud, D. Sarrut, J.-M. Letang. Filtered backprojection proton CT reconstruction along most likely paths. Medical Physics, 2013, 40 (3), pp.031103. ⟨10.1118/1.4789589⟩. ⟨hal-00796949⟩
102 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More