High order finite element simulations for fluid dynamics validated by experimental data from the fda benchmark nozzle model - CNRS - Centre national de la recherche scientifique Accéder directement au contenu
Communication Dans Un Congrès Année : 2017

High order finite element simulations for fluid dynamics validated by experimental data from the fda benchmark nozzle model

Résumé

The objective of the present work is to construct a sound mathematical, numerical and computational framework relevant to blood flow simulations and to assess it through a careful validation against experimental data. We perform simulations of a benchmark proposed by the FDA for fluid flow in an idealized medical device, under different flow regimes. The results are evaluated using metrics proposed in the literature and the findings are in very good agreement with the validation experiment.
Fichier principal
Vignette du fichier
chabannes_prudhomme_szopos_tarabay.pdf (229.97 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01429685 , version 1 (09-01-2017)
hal-01429685 , version 2 (09-01-2017)

Licence

Paternité - Pas de modifications

Identifiants

Citer

Vincent Chabannes, Christophe Prud'Homme, Marcela Szopos, Ranine Tarabay. High order finite element simulations for fluid dynamics validated by experimental data from the fda benchmark nozzle model. 5th International Conference on Computational and Mathematical Biomedical Engineering - CMBE2017, Apr 2017, Pittsburgh, PA, United States. ⟨hal-01429685v2⟩