Decomposing a prior-CT-based cone-beam CT projection correction algorithm into scatter and beam hardening components

Résumé

A scatter correction algorithm (SCA) for cone beam CT (CBCT) projections, making use of prior information obtained by deformable image registration of CT to CBCT, has recently been proposed and tested for particle therapy dose calculation. The SCA relies on subtraction of digitally reconstructed radiographs (DRR) from scaled measured projections and smoothing operations, followed by a subtraction correction and reconstruction. In this note, we compared the SCA's correction to one based on a Monte Carlo simulation of the scatter, and a separate beam hardening correction. Agreement better than 3% between the two approaches was obtained when comparing corrected log-transformed projections.
Fichier principal
Vignette du fichier
1-s2.0-S2405631617300064-main.pdf (352.04 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

hal-01743232 , version 1 (26-03-2018)

Identifiants

Citer

Christoph Zöllner, Simon Rit, Christopher Kurz, Gloria Vilches-Freixas, Florian Kamp, et al.. Decomposing a prior-CT-based cone-beam CT projection correction algorithm into scatter and beam hardening components. Physics and Imaging in Radiation Oncology, 2017, 3, pp.49 - 52. ⟨10.1016/j.phro.2017.09.002⟩. ⟨hal-01743232⟩
179 Consultations
189 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More