Clearsight (TM) use for haemodynamic monitoring during the third trimester of pregnancy - a validation study - CNRS - Centre national de la recherche scientifique Accéder directement au contenu
Article Dans Une Revue International Journal of Obstetric Anesthesia Année : 2018

Clearsight (TM) use for haemodynamic monitoring during the third trimester of pregnancy - a validation study

A. Hili
  • Fonction : Auteur
N. Resseguier
C. Kelway
  • Fonction : Auteur
M. Haddam
  • Fonction : Auteur
A. Bourgoin
  • Fonction : Auteur
X. Carcopino

Résumé

Background: We assessed the validity of Clearsight (TM) as a non-invasive cardiac output and stroke volume monitoring device, comparing it with transthoracic echocardiography measurements during the third trimester of pregnancy. Methods: Measurements obtained from Clearsight (TM) were compared with those from echocardiography as the gold standard. The precision and accuracy of the Clearsight (TM) was measured using the Bland and Altman method. Clinical agreement with echocardiography was assessed using the agreement tolerability index. Results: Measurements were recorded from 44 pregnant women with a median [IQR range] gestational age of 33 [30-37] weeks. We found that Clearsight (TM) measurements presented a systematic overestimation of cardiac output, with mean bias [CI 95%] of 2.7 [2.3-3.0] L/min, with limits of agreement of -0.1 to 5.4 L/min. It overestimated stroke volume, with a bias of 29.5 [25.0-33.4] mL and a limit of agreement of -1.6 to 60.1 mL. In addition, the analysis of cardiac output showed a percentage of error of 41% and intra-class correlation [CI 95%] of 0.37 [0.17 to 0.53, P <0.001]. For stroke volume, the percentage of error was 40% and intra-class correlation 0.16 [-0.1 to 0.34; P=0.27]. We found that agreement tolerability index scores were unacceptable. We evaluated the ability of the device to track changes in cardiac output by inducing a left lateral decubitus position, but the analysis was inconclusive. Conclusion: The agreement between Clearsight (TM) and the echocardiography measurements of cardiac output and stroke volume were not within an acceptable range in the third trimester of pregnancy. (C) 2018 Elsevier Ltd. All rights reserved.
Fichier non déposé

Dates et versions

hal-01970231 , version 1 (04-01-2019)

Identifiants

Citer

G. Duclos, A. Hili, N. Resseguier, C. Kelway, M. Haddam, et al.. Clearsight (TM) use for haemodynamic monitoring during the third trimester of pregnancy - a validation study. International Journal of Obstetric Anesthesia, 2018, 36, pp.85-95. ⟨10.1016/j.ijoa.2018.04.009⟩. ⟨hal-01970231⟩
131 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More