Light response and efficiency calibration of EJ309 liquid scintillator between 3 and 45 MeV - CNRS - Centre national de la recherche scientifique Accéder directement au contenu
Article Dans Une Revue Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment Année : 2019

Light response and efficiency calibration of EJ309 liquid scintillator between 3 and 45 MeV

Résumé

The Neutrons For Science (NFS) facility will provide intense neutron beams in the energy range between 1 MeV and 40 MeV. The neutron monitoring will be ensured by a detector consisting of a glass cell containing EJ309 liquid scintillator coupled to a photomultiplier. The goal of this study was the characterization of this monitor. Two experiments were performed to measure the light response of the scintillator from 3.6 MeV to 45 MeV as well as the detection efficiency between 3.6 MeV and 15.9 MeV. The obtained light response was used to adapt the simulation program SCINFUL to EJ309 scintillator. SCINFUL was then used to reproduce within 5 percent uncertainty the detector efficiency determined experimentally.
Fichier non déposé

Dates et versions

hal-02188096 , version 1 (18-07-2019)

Identifiants

Citer

M.O. Frégeau, M. Delarue, A.-M. Frelin-Labalme, B. Laurent, X. Ledoux, et al.. Light response and efficiency calibration of EJ309 liquid scintillator between 3 and 45 MeV. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2019, 941, pp.162301. ⟨10.1016/j.nima.2019.06.042⟩. ⟨hal-02188096⟩