A strategy to reduce the sensitivity of inhomogeneous Magnetization Transfer (ihMT) imaging to radiofrequency transmit field variations at 3 T

Thomas Troalen

Résumé

Purpose To minimize the sensitivity of inhomogeneous magnetization transfer gradient-echo (ihMT-GRE) imaging to radiofrequency (RF) transmit field (urn:x-wiley:07403194:media:mrm29055:mrm29055-math-0003) inhomogeneities at 3 T. Methods The ihMT-GRE sequence was optimized by varying the concentration of the RF saturation energy over time, obtained by increasing the saturation pulse power while extending the sequence repetition time (TR). Different protocols were tested using numerical simulations and human in vivo experiments in the brain white matter (WM) of healthy subjects at 3 T. The sensitivity of the ihMT ratio (ihMTR) to urn:x-wiley:07403194:media:mrm29055:mrm29055-math-0004 variations was investigated by comparing measurements obtained at nominal transmitter adjustments and following a 20% global urn:x-wiley:07403194:media:mrm29055:mrm29055-math-0005 drop. The resulting relative variations (δihMTR) were evaluated voxelwise as a function of the local urn:x-wiley:07403194:media:mrm29055:mrm29055-math-0006 distribution. The reproducibility of the protocol providing minimal urn:x-wiley:07403194:media:mrm29055:mrm29055-math-0007 bias was assessed in a test-retest experiment. Results In line with simulations, ihMT-GRE experiments conducted at high concentration of the RF energy over time demonstrated strong reduction of the urn:x-wiley:07403194:media:mrm29055:mrm29055-math-0008 inhomogeneity effects in the human WM. Under the optimal conditions of 350-ms TR and 3-µT root mean square (RMS) saturation power, 73% of all WM voxels presented δihMTR below 10%. Reproducibility analysis yielded a close-to-zero systematic bias (ΔihMTR = −0.081%) and a high correlation (ρ² = 0.977) between test and retest experiments. Conclusion Concentrating RF saturation energy in ihMT-GRE sequences mitigates the sensitivity of the ihMTR to urn:x-wiley:07403194:media:mrm29055:mrm29055-math-0009 variations and allows for clinical-ready ihMT imaging at 3 T. This feature is of particular interest for high and ultra-high field applications.

Domaines

Imagerie
Fichier principal
Vignette du fichier
MRM-hal.pdf (2.12 Mo) Télécharger le fichier

Dates et versions

hal-03429467 , version 1 (15-11-2021)

Identifiants

Citer

Lucas Soustelle, Thomas Troalen, Andreea Hertanu, Samira Mchinda, Jean‐philippe Ranjeva, et al.. A strategy to reduce the sensitivity of inhomogeneous Magnetization Transfer (ihMT) imaging to radiofrequency transmit field variations at 3 T. Magnetic Resonance in Medicine, 2021, ⟨10.1002/mrm.29055⟩. ⟨hal-03429467⟩
43 Consultations
145 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More