Results from Verification of Reference Irradiance and Radiance Sources Laboratory Calibration Experiment Campaign

Agnieszka Białek
  • Fonction : Auteur
  • PersonId : 1122730
Emma Woolliams
  • Fonction : Auteur
  • PersonId : 1140309
Joel Kuusk
  • Fonction : Auteur
  • PersonId : 1140311
Ilmar Ansko
  • Fonction : Auteur
  • PersonId : 1140312
Viktor Vabson
  • Fonction : Auteur
  • PersonId : 1140313
Christopher Maclellan
  • Fonction : Auteur
  • PersonId : 1140315
Sabine Marty
  • Fonction : Auteur
  • PersonId : 1140316
William Servantes
  • Fonction : Auteur
  • PersonId : 1122741
Sarah Taylor
  • Fonction : Auteur
  • PersonId : 1122742
Ronnie van Dommelen
  • Fonction : Auteur
  • PersonId : 1122743
Andrew Barnard
  • Fonction : Auteur
  • PersonId : 1122744
Nigel Fox
  • Fonction : Auteur
  • PersonId : 1140318
Riho Vendt
  • Fonction : Auteur
  • PersonId : 1140319
Craig Donlon
  • Fonction : Auteur
  • PersonId : 1098318
Tânia Casal
  • Fonction : Auteur
  • PersonId : 1122745

Résumé

We present the results from Verification of Reference Irradiance and Radiance Sources Laboratory Calibration Experiment Campaign. Ten international laboratories took part in the measurements. The spectral irradiance comparison included the measurements of the 1000 W tungsten halogen filament lamps in the spectral range of 350 nm–900 nm in the pilot laboratory. The radiance comparison took a form of round robin where each participant in turn received two transfer radiometers and did the radiance calibration in their own laboratory. The transfer radiometers have seven spectral bands covering the wavelength range from 400 nm–700 nm. The irradiance comparison results showed an agreement between all lamps within ±1.5%. The radiance comparison results presented higher than expected discrepancies at the level of ±4%. Additional investigation to determine the causes for these discrepancies identified them as a combination of the size-of-source effect and instrument effective field of view that affected some of the results.
Fichier principal
Vignette du fichier
Bialek_etal_RS_2020.pdf (881.12 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-03521261 , version 1 (11-01-2022)

Identifiants

Citer

Agnieszka Białek, Teresa Goodman, Emma Woolliams, Johannes F S Brachmann, Thomas Schwarzmaier, et al.. Results from Verification of Reference Irradiance and Radiance Sources Laboratory Calibration Experiment Campaign. Remote Sensing, 2020, 12 (14), pp.2220. ⟨10.3390/rs12142220⟩. ⟨hal-03521261⟩
18 Consultations
63 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More