Monitoring the Bragg peak location of 73 MeV/u carbon ions by means of prompt $\gamma$-ray measurements - CNRS - Centre national de la recherche scientifique Accéder directement au contenu
Article Dans Une Revue Applied Physics Letters Année : 2008

Monitoring the Bragg peak location of 73 MeV/u carbon ions by means of prompt $\gamma$-ray measurements

Résumé

By means of a time-of-flight technique, we measured the longitudinal profile of prompt $\gamma$-rays emitted by 73 MeV/u $^{13}$C ions irradiating a PMMA target. This technique allowed us to minimize the shielding against neutrons and scattered $\gamma$-rays, and to correlate prompt gamma emission to the ion path. This correlation, together with a high counting rate, paves the way toward real-time monitoring of the longitudinal dose profile during ion therapy treatments. Moreover, the time correlation between the prompt gamma detection and the transverse position of the incident ions measured by a beam monitor can provide real-time 3D control of the irradiation.
Fichier principal
Vignette du fichier
Gamma_prompts_v10.3_preprint_TeX.pdf (174.99 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

in2p3-00315797 , version 1 (01-09-2008)

Identifiants

Citer

E. Testa, M. Bajard, M. Chevallier, D. Dauvergne, F. Le Foulher, et al.. Monitoring the Bragg peak location of 73 MeV/u carbon ions by means of prompt $\gamma$-ray measurements. Applied Physics Letters, 2008, 93, pp.093506. ⟨10.1063/1.2975841⟩. ⟨in2p3-00315797⟩
28 Consultations
41 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More