Reliability of the nitrogen washin-washout technique to assess end-expiratory lung volume at variable PEEP and tidal volumes - CNRS - Centre national de la recherche scientifique Accéder directement au contenu
Article Dans Une Revue Intensive Care Medicine Experimental Année : 2014

Reliability of the nitrogen washin-washout technique to assess end-expiratory lung volume at variable PEEP and tidal volumes

Résumé

Background
End-expiratory lung volume measurement by the nitrogen washin-washout technique (EELVWI-WO) may help titrating positive end-expiratory pressure (PEEP) during acute respiratory distress syndrome (ARDS). Validation of this technique has been previously performed using computed tomography (EELVCT), but at mild PEEP levels, and relatively low fraction of inspired oxygen (FiO2), which may have insufficiently challenged the validity of this technique. The aims of this study were (1) to evaluate the reliability of EELVWI-WO measurements at different PEEP and V T during experimental ARDS and (2) to evaluate trending ability of EELVWI-WO to detect EELV changes over time.
Methods
ARDS was induced in 14 piglets by saline lavage. Optimal PEEP was selected during a decremental PEEP trial, based on best compliance, best EELVWI-WO, or a PEEP-FiO2 table. Eight V T (4 to 20 mL * kg-1) were finally applied at optimal PEEP. EELVWI-WO and EELVCT were determined after ARDS onset, at variable PEEP and V T.
Results
EELVWI-WO underestimated EELVCT with a non-constant linear bias, as it decreased with increasing EELV. Limits of agreement for bias were ±398 mL. Bias between methods was greater at high PEEP, and further increased when high PEEP was combined with low V T. Concordance rate of EELV changes between consecutive measurements was fair (79%). Diagnostic accuracy was good for detection of absolute EELV changes above 200 mL (AUC = 0.79).
Conclusions
The reliability of the WI-WO technique is critically dependent on ventilatory settings, but sufficient to accurately detect EELV change greater than 200 mL.
Fichier principal
Vignette du fichier
2197-425X-2-10.pdf (661.39 Ko) Télécharger le fichier
2197-425X-2-10-S1.DOCX (15.06 Ko) Télécharger le fichier
2197-425X-2-10-S2.DOCX (17.37 Ko) Télécharger le fichier
2197-425X-2-10-S3.DOCX (277.95 Ko) Télécharger le fichier
2197-425X-2-10-S4.DOCX (137.24 Ko) Télécharger le fichier
2197-425X-2-10.xml (75.87 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Format : Autre
Format : Autre
Format : Autre
Format : Autre
Format : Autre
Loading...

Dates et versions

inserm-00976479 , version 1 (09-04-2014)

Identifiants

Citer

Jean-Christophe Richard, Céline Pouzot Pouzot-Névoret, Alfredo Morales Pinzón, Juan Sebastian Torres González, Maciej Orkisz, et al.. Reliability of the nitrogen washin-washout technique to assess end-expiratory lung volume at variable PEEP and tidal volumes. Intensive Care Medicine Experimental, 2014, 2 (1), pp.10. ⟨10.1186/2197-425X-2-10⟩. ⟨inserm-00976479⟩
268 Consultations
345 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More