Design of the detection chain for Athena X-IFU
Résumé
The x-ray integral field unit (X-IFU) instrument is the high-resolution x-ray spectrometer of the ESA Athena x-ray observatory. X-IFU will deliver spectra from 0.2 to 12 keV with a spectral resolution of 2.5 eV up to 7 keV from 5" pixels, with a hexagonal field of view of 5' equivalent diameter. The main sensor array and its associated detection chain is one of the major sub-systems of the X-IFU instrument, and is the main contributor to X-IFU’s performance. CNES (the French Space Agency) is leading the development of X-IFU; additional major partners are NASA-GFSC, SRON, VTT, APC, NIST, and IRAP. This paper updates the B-phase definition of the X-IFU detection chain. The readout is based on time-division multiplexing (TDM). The different sub-components of the detection chain (the main sensor array, the cold electronics stages, and the warm electronics) require global design optimization in order to achieve the best performance. The detection chain’s sensitivity to the EMI/EMC environment requires detailed analysis and implementation of dedicated design solutions. This paper focuses on these aspects while providing an update to the detection-chain design description. © COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.